Chemistry Comp Test Answers – Flashcards
Unlock all answers in this set
Unlock answersquestion
External Calibration |
answer
Set of known concentrations of standards analyzed at the same time as unknowns |
question
Calibration Equation |
answer
y = ax + b |
question
Analytical Equation |
answer
x = (y-a)/b |
question
Standard Addition |
answer
Not used as much Known amounts of standard added to each sample Uses lots of unknown sample |
question
Internal Standardization |
answer
Commonly used in toxicology Useful where sample prep leads to loss of analytes Cmpd added to unknown, distinguish IS from analyte of interest Response ratios of analyte to a fixed quantity of IS plotted for multiple calibrators |
question
Resolution |
answer
R = 2*(t2-t1)/(Wb2+Wb1) Recommended R ~2.0 |
question
Theoretical Plates |
answer
Neff = 5.54(t'R)^2/(W1/2)^2 Efficiency of column |
question
Carrier Gas Effects |
answer
Resolution and efficiency Nitrogen- better resolution Helium and Hydrogen- give up resolution, increase ability to have shorter run times |
question
Van Deempter Equation |
answer
HETP = A + B/u + C*u |
question
Van Deempter A |
answer
Eddy diffusion term Particle size distribution Not as in important in HPLC |
question
Van Deempter B |
answer
Longitudinal diffusion term Proportional to time on column Density of mobile phase affects Increasing temperature causes molecules to diffuse more and move faster, B increases |
question
Van Deempter C |
answer
Mass transfer term Due to inadequate time for equilibrium between phases Faster flow = less time |
question
Splitless Inlet System |
answer
Want maximum sensitivity Use of small sample amount |
question
GC Isothermal Elution |
answer
Not useful for complex samples with many compounds over wide BP range Early cmpds bunched, heavier cmpds have long elution times with low peak heights |
question
GC Temperature Programming |
answer
Use for samples with wide range of volatility, boiling points, or molecular weights Uses low column temperatures to better separate early cmpds, increasing temps to speed the elution of more retained cmpds (cuts analysis time) |
question
GC Thermal Conductivity Detector (TCD) |
answer
Most "universal", not very sensitive (1 ng) Best for fixed gases: N2, O2, CO2, CO |
question
GC Flame Ionization Detector (FID) |
answer
"Universal" for most organic compounds (10-100 pg) Not sensitive to fixed gases or water Requires H2 and "Zero Air" for flame Most commonly used (being replaced by MS) Drug quantitation |
question
GC Nitrogen-Phosphorus Detector (NPD) Alkali-Flame Ionization Detector (AFID) Thermoionic |
answer
Selective for N or P containing cmpds Was often used in drug and tox labs |
question
GC Nitrogen-Phosphorus Detector (NPD) Alkali-Flame Ionization Detector (AFID) Thermoionic |
answer
Selective for N or P containing cmpds (10 pg) Drug analysis Being replaced by MS detectors |
question
GC Electron Capture Detector (ECD) |
answer
Selective for halogen, nitro, or aromatic cmpds (50 fg) Not used often Contains radioactive source (63Ni) |
question
GC Infrared Detector (GC-IRD) |
answer
Universal for most organic cmpds Not as sensitive as other detectors Good for ID of isomers of cmpds where mass spectra are similar |
question
GC-MS |
answer
Most widely used instrumental technique in chemistry lab Samples not good- explosives (thermally unstable), low volatility materials, inorganic salts, fibers |
question
Mass Spectrometry- GC Detector |
answer
Universal and specific (1 ng and 1 pg) TIC- all ions in range EIC or Selective Ion Monitor- selective |
question
MS Ion Source- Electron Impact (EI) |
answer
Most common Ions stream off tungsten filament to target Hit and knock off electron to become positive |
question
MS Ion Source- Chemical Ionization (CI) |
answer
Creates positive ion Useful if a weak molecular ion under EI conditions If cmpd fragments a lot |
question
MS Ion Source- Negative Chemical Ionization (NCI) |
answer
Creates negative ion Useful for explosives |
question
MS Ion Source- Electrospray Ionization |
answer
Used with HPLC Carrier gas Droplets have a positive charges and accelerated across gap with potential |
question
MS Ion Source- MALDI |
answer
Matrix Assisted Laser Desorption Ionization Used for large molecules Pulse with photons of laser energy to blast off chunks Ions streamed through TOF analyzer |
question
Mass Analyzer- Quadrapole |
answer
Most common in forensics Poles paired with fixed voltage or alternating current in radio frequency Ions spirals down length of tube Ions with specific mass can only make it down |
question
Mass Analyzer- Ion Trap |
answer
Also frequently used Ions oscillate inside- only right size will pass through Alternating current |
question
Mass Analyzer- Ion Cyclotron Resonance |
answer
FT-ICR |
question
Mass Analyzer- Time of Flight (TOF) |
answer
Often used with MALDI Energy potential becomes the source Light ions move faster Time it takes is proportional to mass Can determine mass of ion up to 3 decimal places |
question
Mass Analyzer- Ion Mobility Spectrometer |
answer
Electrodes accelerate ions Similar to time of flight Used as portable "sniffer" for drugs, explosives, and chemical warfare agents |
question
Mass Analyzer- Tandem MS/MS |
answer
3 quadrapoles Used with liquid chromatography to screen tox samples |
question
Total Ion Chromatogram (TIC) |
answer
Sum of intensities of all ion fragments |
question
Extracted Ion Chromatogram (EIC) |
answer
Generally "post run" process Can go back until you find what you want Sum of selected ions from each scan Reconstructed Ion Chromatogram Fire debris analysis |
question
Single/Selective Ion Monitoring (SIM) |
answer
Only a few specific ions scanned More sensitive than normal scan More data collection time at each ion Toxicology- at of drug in bodily fluids is small "Pre-run": set up prior to scan, extracts only selected ions Can't go back |
question
Calibration of MS |
answer
Mass calibration done weekly with perfluorinated cmpds |
question
"Tuning the MS" |
answer
Adjustment of ionization, repeller, and accelerator voltages Most MSD have "auto-tune" |
question
Fire Debris Extraction Techniques |
answer
Steam Distillation Simple Headspace Headspace Concentration- Passive (ASTM E1412 Activated Charcoal), Dynamic, SPME (Solid Phase Microextraction) Solvent Extraction |
question
Instrumentation- Analysis of Ignitable Liquids |
answer
GC with FID, GC-MS |
question
Ignitable Liquid Classifications |
answer
ASTM E1618-06 8 classes with 3 potential subclasses for each |
question
Gasoline Recognition |
answer
Blended Product C4-C12 with alkanes, aromatics, cycloparaffins Aromatics most abundant group Xylnes, C3 Castle, C4 Gang of four |
question
EIC Chromatogram Plots- Fire Debris |
answer
Alkanes Cycloalkanes Aromatics Napthenics Indanes |
question
Fire Debris Analysis- Class determinations |
answer
[image] |
question
Spot Tests |
answer
[image] |
question
Scott Test |
answer
Cocaine salts/Base reagent= positive result requires all 3 |
question
Mandelin Test |
answer
Green for amphetamine and methamphetamine |
question
Confirmation Tests for Drugs- Categories of Analytical Techniques |
answer
1 A and 1 B 2 Bs and 1 C [image] |
question
HPLC vs. GC |
answer
More suitable for thermally unstable cmpds- explosives Samples must be particulate free |
question
HPLC Injector |
answer
Precise volume 6 port loop injector often used Fixed volume most reproducible |
question
HPLC Normal Phase |
answer
Non-polar mobile phase with polar stationary phase |
question
HPLC Reverse Phase |
answer
More polar mobile phase with stationary phase Most commonly used |
question
HPLC Isocratic Solvent Selection |
answer
One solvent mixture throughout entire analysis Simple mixtures or polymer analysis |
question
HPLC Gradient Elution |
answer
Changing from one solvent mixture to another over time during the run Analogous to temperature programming in GC Enhances separation of early eluting components while reducing elution time for later components |
question
HPLC UV/Visible Absorption Detectors |
answer
Fixed wavelength Variable wavelength Diode array detection (multiwavelength) |
question
Diode Array Detector |
answer
Multiwavelength Needs to absorb in UV or visible range Lots of double bonds, aromatic character Cmpds with saturated hydrocarbons don't work (fire debris) Light absorbed and then split into specific wavelengths |
question
HPLC Fluorescence Detector |
answer
More sensitive than absorbance detectors Cmpd must fluoresce Used for LSD |
question
HPLC Infrared Absorption Detector |
answer
Solvent interference restricts widespread application |
question
HPLC Refractive Index Detector (RI) |
answer
Least sensitive Best for non UV absorbers (sugars) Not suitable for gradient elution Reference side and sample size |
question
HPLC Mass Spectrometry Detector |
answer
Becoming 2nd most common Very sensitive Solvent removal Allows for better identification of peak |
question
HPLC Forensic Applications |
answer
Drug tox- polar cmpds, metabolites, enantiomeric separation Explosives- thermally unstable, good UV absorbers Questioned documents- analysis of inks, dyes Polymer analysis- size exclusion chromatography, fibers, plastics, adhesives |
question
UV/Vis Spectroscopy Cmpd Classes |
answer
Aromatic compounds- controlled substances, metabolites, organic dyes and pigments, some explosives Nitro compounds Transition metal compounds |
question
IR Detector- Deuterated Triglycine Sulfate (DTGS) |
answer
Most widely used detector in IR Wide sensitivity range and quick response Photoelectric transducer |
question
IR Detector- Mercury/Cadmium Telluride (MCT) |
answer
Photoconducting transducer Narrower sensitivity range, but more sensitive Needs liquid nitrogen cooling to reduce thermal noise |
question
IR Diffuse Reflectance |
answer
Rough surface scatters light in all directions Sample mixed with non-absorbing material (non-compressed KBr) or on surface of silica carbide paper |
question
IR Attenuated Total Reflectance |
answer
Internal Reflectance- ATR commonly used Good for opaque or other hard to hand samples (pastes, adhesives) Doesn't require any big prep work |
question
IR Forensic Applications |
answer
Controlled substances (crack cocaine vs. cocaine salt) Automobile make, model, year from paint fragments Polymers (fiber analysis, tapes, adhesives) Infrared microscope- trace analysis, fibers, paint |
question
IR Frequency Table |
answer
[image] |
question
Raman with FTIR Sampling |
answer
Aqueous solutions and glass sample containers can be used Minimal sample prep |
question
Raman with FTIR- Fluorescence |
answer
Much greater than Raman intensity Broad background with superimposed Raman spectrum (hill with grass on top) "Photobleaching" sometimes effective |
question
Raman with FTIR- Laser Power |
answer
Must be adjusted to prevent burning of trace sample- especially dark colored materials (black tar heroin, smokeless powder) |
question
Raman with FTIR Forensic Applications |
answer
Seized drug analysis Scan through baggies, glass vials Minimal sample prep Portable units available to take to clandestine lab scenes Explosives analysis Paints Fibers and other polymers Inks |
question
Paint Layers |
answer
Clearcoat Effect coat Basecoat Surfacer Electrocoat Primer |
question
Analysis Methods of Paint |
answer
Solubility Pyrolysis GC FTIR Microscopy |
question
TLC Normal Phase |
answer
Polar silica stationary phase, relatively non-polar mobile phase |
question
TLC UV Fluorescence |
answer
Most common 254 nm and 365 nm Analyte absorbs UV- dark spot, bright bkd No bkg fluorescence with 365 nm, but analyte may fluoresce |
question
TLC Universal Reagents |
answer
Iodine vapor (double bonds)- reversible Conc. sulfuric acid w/heating |
question
TLC Group Specific Reagents |
answer
Ninhydrin- amine groups Iodoplatinate- drugs Fast blue B- cannabinoids Ehrlich- LSD Diphenylcarbizone/HgSO4- barbiturates |
question
TLC Forensic Applications |
answer
Drug screening Ink analysis Explosives screening |
question
Ink Analysis Forensic Applications |
answer
Counterfeit currency Authentication of historical documents Written materials relative to another crime Ink stain comparison |
question
Document Accelerated Aging Approaches |
answer
UV Accelerated Aging- dye partially degraded, solvent still present Thermal Accelerated Aging- solvent gone, dye intact |
question
Age Dating Inks Analysis |
answer
Dye composition changes: HPLC-DAD and laser desorption MS Changes in ink solubility with time Solvent changes (evaporation) Changes in dye composition |
question
High Explosives |
answer
TNT PETN ANFO Nitroglycerin HMX |
question
Low Explosives |
answer
Smokeless Powders Black Powder (Pyrodex) Pyrotechniques |
question
Smokeless Powder Shapes |
answer
Ball Disk Rod or tube Characteristic of manufacturer and brand |
question
Explosive Analysis (Post Blast) |
answer
GC/MS primary HPLC and CE SEC TLC |
question
Explosive Device Detection (Pre Blast) |
answer
Ion mobility spectroscopy (IMS) Canine X-ray imaging |
question
Metals Analysis- Flame Atomic Absorption |
answer
Single element analysis Least sensitive method Best for alkali metals (Li, Na, K) |
question
Metals Analysis- Graphite Furnace AA Electrothermal or Flameless AA |
answer
Single element Good to ppb levels for some metals Very small amts of sample |
question
Metals Analysis- Inductively Coupled Plasma Optical Emission Spectroscopy (ICP-OES) |
answer
Simultaneous elements Good to sub ppb levels (most sensitive) |
question
Metals Analysis- X-Ray Fluorescence |
answer
Very little sample prep |
question
Metals Analysis- ICP-MS |
answer
Best for toxicology Simultaneous multi-element method Very sensitive Capable of isotopic composition of element |
question
Metals Analysis- Neutron Activation Analysis (NAA) |
answer
Little or no pretreatment Needs nuclear reactor as source of neutrons Non-routine analysis Specialized |
question
Metals Analysis Forensic Applications |
answer
Bullet lead comparison GSR Explosives Trace elements in glass (Laser Ablation ICP-MS) Metallic poisons |
question
NIDA-10 Drugs (*NIDA-5 Drugs) |
answer
Amphetamines* Barbiturates Benzodiazapines Cannabinoids* Cocaine* LSD Methadone Opiates* Phencyclidine(PCP)* Propoxyphene |
question
Strongly Acidic Drugs |
answer
Salicylic acid Pheylbutazone 2,4-Dinitrophenol |
question
Weakly Acidic Drugs |
answer
Acetominophen Phenobarbital Chlorothiazide |
question
Amphoteric Basic Drugs |
answer
Morphine Gabapentin Benzoylecgonine |
question
Strongly Basic Drugs |
answer
Amphetamine Chlorcyclizine |
question
Plasma |
answer
Whole blood centrifuged to remove cellular material, clotting factors still present |
question
Serum |
answer
Similar to plasma but does not have clotting factors |
question
Denaturation of Proteins |
answer
Elevated temperature pH Organic solvents Enzymatic breakdown of protein or conjugate |
question
Tox GCMS Internal Standard |
answer
Deuterated drug or metabolite is best |
question
Tox GCMS Confirmation |
answer
SIM usually required for sensitivity Target ion plus at least 2 confirmatory ions |
question
Tox LCMS/MS |
answer
Becoming preferred method Minimal sample prep |
question
Tox Fractionation Scheme |
answer
[image] |
question
Salt vs. Free Base IR |
answer
OH band generally broader and more absorbing than free base |
question
Fingerprint Region in IR Spectrum |
answer
Below 1200 cm-1 |
question
Band found at 2350 cm-1 in a background KBr pellet |
answer
Carbon dioxide in the atmosphere |
question
Primary and Secondary Explosives |
answer
Primary- shock sensitive Secondary- requires detonator |
question
Vibrational Spectroscopies- Raman vs. IR |
answer
Raman spectrum- inelastic scattering IR- absorbance of light photons |
question
BAC Analyzed By: |
answer
Headspace sample GC with FID N-Proponal as IS |