Anesthesia Machine: High, Intermediate & Low Pressure Systems

Unlock all answers in this set

Unlock answers
question
What are the 3 parts of the pneumatic part of the anesthesia machine?
answer
high, intermediate and low pressure systems
question
high-pressure system
answer
receives gases from cylinders at high, variable pressures and reduces those pressures to lower, more constant pressure suitable for use in the machine
question
hanger yoke
answer
orients and supports the cylinder, provided a gas-tight seal and ensures a unidirectional gas flow
question
body of the yoke
answer
threaded into the frame of the machine and provides support for the cylinders
question
retaining screw also known as
answer
also known as the clamping device or retaining bar
question
Purpose of retaining screw
answer
threaded into the distal end of the yoke and when tightened, presses the cylinder valve outlet against the washer and nipple so a gas-tight seal is achieved
question
nipple
answer
part of the yoke through which the gas enters the machine
question
why is it important that cylinder valves and yokes not be contaminated
answer
fire hazard
question
pressure regulator
answer
reduces the high and variable pressure found in a cylinder to a lower, more constant pressure suitable for use in an anesthesia machine. Can be direct or indirect acting
question
pressure
answer
force acting against an area
question
intermediate pressure system
answer
receives gases from the pressure regulator or the pipeline inlet to the anesthesia machine
question
what are the pneumatic components of the intermediate pressure system
answer
pneumatic part of the master switch, pipeline inlet connections, pipeline pressure indicators, piping, the gas power outlet, oxygen pressure failure devices, the oxygen flush, flow control valves
question
piping
answer
used to connect components inside the machine and must be able to withstand 4x the intended service pressure without rupturing
question
oxygen failure safety device
answer
whenever the oxygen supply pressure falls below the minimum, the delivered oxygen concentration will not decrease below 19%
question
where is the low pressure system located
answer
downstream of the flow control devices
question
what are the components of the low pressure system
answer
flowmeters, hypoxia prevention safety devices, unidirectional valves, pressure relief devices, common gas outlet, vaporizers
question
flowmeter
answer
indicate the rate of flow of a gas passing through them
question
what principle is measuring gas flow in a mechanical meter based on
answer
flow past resistance is proportional to pressure
question
rate of flow through the tube depends on
answer
pressure drop against constriction, size of the annular opening, physical properties of the gas
question
what does the flowmeter assembly consist of
answer
the tube through which the gas flows, the indicator, a stop at the top of the tube and the scale that indicates the flow
question
where is the greatest accuracy of the flowmeter
answer
middle half of the tube
question
common (fresh) gas outlet
answer
receives all of the gases and vapors from the machine and delivers the mixture to the breathing system
question
pin index safety system
answer
has specific locations for pins and holes for each gas
question
when placing a cylinder in a yoke
answer
you must fully retract the retaining screw and pin index safety system must be mated
question
if a flowmeter tube is broken
answer
the tube, the indicator and scale must be replaced as a unit
question
at low flows, flow is a function of
answer
the viscosity of the gas
question
viscosity of the gas is more important with
answer
a longer and narrower constriction
question
with high flows, flow is a function of
answer
the density of the gas
question
what should the oxygen flowmeter be next to
answer
the manifold outlet
question
a leak in the flowmeter that is farthest from the manifold outlet may result in
answer
a loss of gas from the middle flowmeter
question
why should gas cylinder valves on the anesthesia machine with a pipeline supply by kept closed
answer
so that pipeline pressure failure will be recognized early, to prevent gas from flowing retrograde into the pipeline system, to prevent pressure alterations in the pipeline from drawing down the cylinder contents