Outcome 2.2 – Oxygen Delivery Systems – Flashcards

Unlock all answers in this set

Unlock answers
question
High flow system vs low flow system
answer
high:device that meets al of the patient's inspiratory flow demands low: does not meet patient's insp. demands. NOTE: nothing to do with how much FiO2 delivered by device
question
FiO2 depends on:
answer
reservoir size patient's ventilator system oxygen flow
question
Low flow system: pros and cons
answer
pro: convenient, easy to set up. can deliver high and low FiO2. cons: FiO2 variable.
question
Low flow system is suitable when
answer
Vt (tidal volume) is 300-700mL RR < 25 Ventilatory patter regular and consistent consistent and predictable FiO2 not required
question
To determine patient's high flow requirements you need to know some of these:
answer
Minute Volume: (Vt)(RR) Cycle time: start of inspiration to end of expiration Inspiratory time Expiratory time Inspiratory flow
question
Pros and Cons of high flow systems
answer
Pros: high and low FiO2, constant and predictable FiO2, depending on system can control temperature, humidity Cons: set up time consuming, requires regular monitoring, usually more expensive. Suitable anytime consistent and predictable FiO2
question
Low Flow systems: Nasal Cannula
answer
FiO2's: 0.24-0.44 Flow: up to 6 Lpm Application: cannula sits inside nares, while tubing rests over top of ears Pros: uses upper airway as gas reservoir, commonly used Cons: uncomfortable at higher flows, due to dryness
question
Low Flow systems: Nasal Catheter
answer
FiO2: 0.24-0.44 Flow: 1-6Lpm application: inserted through nasal passage until the tip sits behind uvula. pros: none cons: uncomfortable, need to be changed regularly
question
Low Flow systems: High flow nasal cannula
answer
like regular NC, but tubing is a wider bore so flows can be higher and FiO2 delivery is higher. flows up to 15 Lpm replaces partial rebreather mask for FiO2 delivery
question
Low Flow systems: simple mask
answer
FiO2's: 0.4-0.6 flow: 6-10 Lpm application: mask covers mouth and nose, adjustable metal tab covering bridge of nose, flexible straps behind bask
question
Simple Mask, pros and cons
answer
pros:higher FiO2's possible because mask acts as additional oxygen reservoir, can attach to a nebulizer to deliver medications, humidity cons: patient unable to eat, speech may be muffled, flow must be >5Lpm to avoid dangerous CO2 buildup
question
Low Flow systems: partial rebreathing mask
answer
FiO2's: 0.6-0.8+ Flow: is set so bag does not collapse during inspiration. in practice this is 6-15Lpm, at minimum of 6Lpm application: same as simple mask
question
Low Flow systems: non rebreathing mask
answer
FiO2's: 0.8+ flow: as with partial rebreather, flow is set to keep bag inflated application: comes with 3 valves, be sure to remove one for safety. exhaled air does not go into bag. valves on mask to minimize entrainment of room air on inspiration
question
Low Flow systems: Trans Tracheal Catheter
answer
used rarely, almost always in emergency situation placed either through cricoid cartilage or between the second and third tracheal ring uses upper and lower airway as reservoir.
question
High Flow systems: air entrainment mask
answer
•A source gas (in this case oxygen) is diluted by air via jet to produce the desired FiO2 •If we think back to Bernouilli's principle we know that as gas velocity increases, lateral pressure decreases and this allows atmospheric gas to rush in. •An air entrainment mask uses this principle, by altering jet size, and the size of the entrainment port we can control FiO2 and total flow to the patient
question
High Flow systems: optiflow; description
answer
heated, humidified high flow oxygen therapy. wide bore nasal prongs to deliver flow to patient. capable of 0-60 Lpm as either high or low system. theoretically provides PEEP to patient. Two *separate* components: blender to select FiO2; flowmeter to select flow delivery to patient.
question
High Flow systems: optiflow; initiation
answer
start at 35-40 Lpm, based on patient RR. flow can be titrated up or down based on the work of breathing. more flow may help alleviate some work of breathing. start FiO2 at 1.0 then quickly wean until you reach your target saturation. set humidifier to invasive mode (37C)
question
High Flow systems: optiflow; weaning
answer
start by weaning FiO2 as long as you are maintaining your target O2 saturation. once FiO2 below .4, start weaning the flow as long as work of breathing doesn't deteriorate. general rule: consider an FiO2 of .35 and a flow of 20Lpm to be roughly equivalent to 6Lpm via traditional nasal cannula
question
High Flow systems: T piece for tracheotomy patients
answer
similar to optiflow, humidified high flow system. T attaches directly to tracheostomy tube. ensure you have pressure release valve somewhere in system.
question
Pulse Oximetry. tool used to assess:
answer
the need for oxygen therapy effectiveness of our current oxygen therapy
question
Pulse Oximetry, method of operation
answer
oxygenated hemoglobin absorbs different wavelengths of light than non oxygenated. calculate relative concentrations of oxy and de-oxy hemoglobin based on their different reactions of light
question
Pulse Oximetry; features
answer
non invasive measures oxygen saturation of hemoglobin can also measure heart rate functions by measures of photospectrometry principle of light absorption
question
Spectrophotometry in pulse oximetry
answer
used to determine oxy and deoxy hemoglobin only shines diff wavelengths of light through blood red (660nm) deoxygenated HgB infrared (940nm) oxygenated HgB also relies on pulsatile arterial flow; during systole the increased volume of blood increases the light absorbed during diastole the lower volume of blood decreases absorption
question
Pulse oximetry; principle of operation
answer
2 LED's transmit the red, infrared light through the vascular bed a photodiode measures light which passes through the vascular bed. (finger, toe, hand/foot, earlobe, forehead) Ratios between the amplitude of the two plethysmographic waveforms are converted into an arterial oxygen saturation
question
Accuracy of pulse oximeter
answer
check pulse index (around 1.0) check actual pulse, for correlation to sensor check waveform on monitor
question
oxyhemoglobin dissociation curve
answer
left shift - hemoglobin have high affinity for oxygen higher pH, Lower DPG, lower temp right shift - hemoglobin have low affinity for oxygen. lower pH, higher DPG, higher temp
question
pulse oximeter device limitations
answer
if dyshemoglobins are present in patients, may lead to inaccuracies: carboxyhemoglobin -ex smoke inhalation. binds strongly to hemoglobin. creates bright cherry red color to arterial blood methemoglobin -iron molecule in hemoglobin oxidizes. turns blood muddy brown. dyes and pigments: intravascular, nail polish, deeply pigmented skin
question
pulse oximeter device limitations con't
answer
perfusion: ex cold room motion artifact high intensity light cannot read hyperoxemia diminished accuracy of <80%
question
Lambert-Beers Law
answer
•Used when converting light absorbance into concentrations •Measured absorbance for a single substance is directly proportional to its concentration and the length of the light path through the sample
question
Lambert-Beers Law Formula and coefficients
answer
A = e × c × l A = absorbance e = extinction coefficient (how light fades through a substance) c = concentration l = length of light path
question
Different Hemoglobins (6)
answer
HHb O2Hb COHb MetHb SHb Bilirubin
question
Dyshemoglobins: Carboxyhemoglobin COHb
answer
In COHb the binding sites are occupied by carbon-monoxide. • Cherry red discoloration of blood • Caused by inhalation from: Tobacco fumes, Exhaust gases, Fires • Therapy: Maintenance of adequate ventilation, high FIO2, hyperbaric oxygen therapy • The Oxygen Dissociation Curve is shifted to the left
question
Dyshemoglobins: Methemoglobin - MetHb
answer
In MetHb the iron atom in one to four hemes of the hemoglobin tetramer is in ferri state. The hemes in this iron-state do not bind O2 but bind water. • Brownish discoloration of blood caused by oxidizing chemicals or drugs • Caused by chemicals: Aniline dyes - Nitrobenzene - Nitrotoluene -Naphthalene - Phenylhydrazine and medicine: Nitrites - Para-aminosalicylic acid - Prilocaine - Benzocaine - Lidocaine - Nitric Oxide • Therapy: Methylene blue, Ascorbic Acid • The Oxygen Dissociation Curve is shifted to the left
question
Dyshemoglobins: Sulfhemoglobin - SulfHb
answer
one or two hemes of the hemoglobin tetramer are sulfated. The compound is very stable and has reduced affinity for oxygen. • Greenish discoloration of blood • Caused by medicine containing sulphur • Therapy: No therapy available, once formed sulfhemoglobinemia will persist for the life of the red cell (~ 120 days) and cause cyanosis for several months after toxic exposure • The Oxygen Dissociation Curve is shifted to the right
Get an explanation on any task
Get unstuck with the help of our AI assistant in seconds
New